Medical Slide Holder FH-G1 User’s Manual
Adapter für medizinische Präparate FH-G1 Bedienungsanleitung
Porte-Objet Médical FH-G1 Manuel de l’utilisateur
Portaobjetos para Fines Médicos FH-G1 Manual del Usuario
Objectglashouder FH-G1 Gebruikshandleiding
Porta Vetrini Medici FH-G1 Guida per l’ utente
6MM51570
Printed in Japan
/ Figure C / Abbildung C / Figure C / Figura C /
Figuur C / Figura C
/ Figure A / Abbildung A / Figure A / Figura A /
Figuur A / Figura A
/ Figure B / Abbildung B / Figure B / Figura B /
Figuur B / Figura B
MA-20 (S)
FH-G1
4 5
Vielen Dank für Ihr Vertrauen in Nikon-Tech-
nologie.
Mit Hilfe des Adapters FH-G1 können Sie
mikroskopische Durchlichtpräparate mit den
Nikon-Kleinbild-/APS-Filmscannern SUPER
COOLSCAN 4000 ED (4000 ED) und
COOLSCAN IV ED, LS-2000, COOLSCAN III
digitalisieren.
Um einen ordnungsgemäßen Betrieb des Nikon-
Scanners mit dem Präparatadapter zu gewähr-
leisten, sollten Sie diese Bedienungsanleitung
und das Handbuch des Scanners aufmerksam
lesen.
Hinweise
Änderungen vorbehalten.
Nikon übernimmt keine Haftung für Beschä-
digungen an Präparaten, die durch die Benut-
zung dieses Adapters entstehen.
Sicherheitshinweise
Bitte lesen Sie diese Bedienungsanleitung auf-
merksam durch, bevor Sie den Präparatadapter
verwenden. Machen Sie diese Anleitung allen
Personen zugänglich, die mit dem Präparat-
adapter arbeiten werden.
Gebrauchshinweise
Achten Sie darauf, daß der Präparatadapter
nicht herunterfällt oder anderen Erschütte-
rungen ausgesetzt ist. Der Adapter oder die
eingelegten Präparate könnten beschädigt
werden, was das Verletzungsrisiko erhöht.
Behandeln Sie den Adapter und die Präpara-
te immer mit größter Sorgfalt.
Präparate, die aufgrund ihrer Stärke nicht
durch die Führung am Anfang des Adapters
passen, können nicht verwendet werden.
Bitte versuchen Sie nicht, Präparate mit Ge-
walt durch die Führung zu drücken. Adapter
und Präparat könnten dabei beschädigt wer-
den, wodurch Verletzungsgefahr entsteht.
Der Präparatadapter kann bei starker me-
chanischer Belastung beschädigt werden.
Lassen Sie ihn nicht fallen, verbiegen Sie ihn
nicht und gehen Sie sorgfältig damit um.
Der Präparatadapter enthält Kunststoffteile,
die sich bei starker Hitzeeinwirkung verfor-
men können. Setzen Sie den Adapter keinen
hohen Temperaturen aus (z.B. durch direkte
Sonneneinstrahlung). Bewahren Sie ihn nicht
in der Nähe von starken Wärmequellen wie
Heizungen auf, und lassen Sie ihn an heißen
Sommertagen nicht im Auto liegen.
Verwenden Sie den Präparatadapter nur mit
den oben aufgeführten Scannern. Bei der
Verwendung mit anderen Scannern können
der Adapter und der Scanner beschädigt
werden.
Anwendung des FH-G1
Präparate
Der Adapter ist für Präparate folgender Abmes-
sungen geeignet:
Objektträger: 26 × 76 mm (B × L), Stärke 0,8
bis 1,5 mm
Deckglas: 25 × 60 mm (B × L), Stärke 0,18
mm oder weniger
Maximale Präparatstärke:
2 mm oder weniger (einschließ-
lich Kitt)
Verwenden Sie keine undurchsichtigen Proben.
Vor dem Scannen
1. Halten Sie den Präparatadapter FH-G1 so,
daß das Nikon-Logo nach oben weist. Schie-
ben Sie das Präparat vorsichtig in die Präparat-
aufnahme des Adapters (Abb. B). Achten Sie
darauf, daß das Deckglas oben liegt.
English Deutsch
* The illustration shows the 4000 ED/
COOLSCAN IV ED with the MA-20 (S).
* Die Abbildung zeigt den 4000 ED/
COOLSCAN IV ED mit dem Diaadapter
MA-20 (S).
* Illustration présentant le 4000 ED/
COOLSCAN IV ED avec MA-20 (S).
* La ilustración muestra el 4000 ED/
COOLSCAN IV ED con el MA-20 (S)
* De illustratie toont de 4000 ED /
COOLSCAN IV ED met de MA-20(S).
* L’immagine mostra il 4000 ED/
COOLSCAN IV con MA-20(S).
2. Richten Sie das Präparat so im Halter aus,
daß es sich innerhalb des Scanbereichs (sie-
he
5
in Abb. A) befindet. Orientieren Sie
sich dabei an den Markierungen (siehe
4
in
Abb. A).
Gebrauchshinweise
Wenn Sie das Präparat in den Adapter
eingesetzt haben, sollten Sie ihn vorsichtig
behandeln. Obwohl das Präparat von den
Montageclips gehalten wird, kann es bei
starken Erschütterungen aus der Halterung
rutschen.
3. Setzen Sie den Diaadapter MA-20 oder
MA-20 (S) in den Scanner ein, und vergewis-
sern Sie sich, daß sowohl der Scanner als
auch der Computer eingeschaltet sind.
4. Schieben Sie den Präparatadapter FH-G1 in
der Richtung des Pfeils – das Nikon-Logo
weist nach oben – in den Diaadapter MA-20
oder MA-20 (S) ein (Abb. C).
Gebrauchshinweise
Achten Sie darauf, daß der Scanner einge-
schaltet ist, wenn Sie den Präparatadapter
in den Scanner einschieben. Sie dürfen den
Scanner nicht einschalten, wenn sich der
Präparatadapter bereits im Scanner befin-
det – andernfalls kann der Scanner keine
Initialisierung durchführen.
Scannen
Nachdem der Präparatadapter in den Scanner
eingeschoben wurde, können Sie die Präparate
wie normale Dias scannen. Nähere Informatio-
nen finden Sie im Handbuch zu Nikon Scan.
Gebrauchshinweise
Bei eingeschalteter Digital ICE cubed (Digi-
tal ICE
3
)-Funktion können Störungen im
Scan sichtbar werden. Schalten Sie die Digi-
tal ICE cubed (Digital ICE
3
)-Funktion aus,
wenn Sie Präparate scannen.
Die untere Kante der Scanmaske ist mögli-
cherweise im Scan zu sehen. Da die untere
Maskenkante tiefer liegt als das Präparat,
befindet sie sich außerhalb des Schärfe-
bereichs. Dabei handelt es sich nicht um eine
Fehlfunktion des Scanners oder des Präparat-
adapters.
Nach dem Scannen
Nachdem Sie Ihr Präparat gescannt haben, neh-
men Sie den Präparatadapter aus dem Scanner
und das Präparat aus dem Adapter. Gehen Sie
dabei in der umgekehrten Reihenfolge wie un-
ter »Vor dem Scannen« beschrieben vor.
Gebrauchshinweise
Möglicherweise haftet der Objektträger et-
was am Halter, wenn Sie versuchen, ihn zu
entnehmen. Schieben Sie ihn einfach vor-
sichtig zur Seite.
Pflege
Reinigen Sie den Adapter, wenn nötig, vorsich-
tig mit einem sauberen Tuch.
Vermeiden Sie jeden Kontakt des Adapters mit
flüchtigen Chemikalien wie Alkohol, Verdünner
oder Benzin. Diese Stoffe greifen die Oberflä-
che des Adapters an.
Technische Daten
Effektiver Scanbereich:
22,9 × 35 mm (der Scanbereich ist
durch die Markierungen gekennzeich-
net)
Abmessungen:
53 × 106 × 7 mm (B × L × H)
Gewicht: ca. 20 g
Thank you for your purchase of an FH-G1
MEDICAL SLIDE HOLDER.
The FH-G1 is designed specifically for scanning
slides glass (for microscope) in Nikon’s SUPER
COOLSCAN 4000 ED (4000 ED) and
COOLSCAN IV ED, LS-2000, COOLSCAN III
35mm/IX240 film scanners.
Using this holder, you can scan slides glass (for
microscope) containing samples of pathological
tissue in much the same way you would ordi-
nary mounted film positives. For trouble-free
operation, be sure that you have read and
understood both the present manual and the
documentation included with your scanner.
Notices
The information contained in this manual is
subject to change without notice.
Nikon does not accept liability for damage to
slides glass (for microscope) resulting from
the use of this product.
For Your Safety
Be sure to read this manual thoroughly before
using the product. After reading, store this
manual where it will be read by all who use the
product.
Precautions for Use
If the product is dropped or subject to other
forms of physical shock, the product or
prepared slide could break, resulting in injury.
Prepared slides of a thickness greater than
the limit imposed by slide insertion guides at
the holder entrance can not be used. At-
tempting to force a prepared slide of greater
thickness into the holder could cause the
holder or slide to break, resulting in injury.
The product may suffer damage if dropped,
twisted, or struck against hard objects.
This product contains reinforced plastic, and
may suffer deformation when subject to
extreme heat. Do not store the product in
direct sunlight, in the vicinity of heat-produc-
ing equipment, in a closed car on a hot day,
or in any other location subject to high
temperatures.
Do not use the product with scanners other
than those listed above, as this could cause
malfunction or result in damage to the scan-
ner or the product.
Using the FH-G1
Slides Glass (for microscope)
This product can be used with slides of the
dimensions given below.
Slide: 26mm (W) × 76mm (L), 0.8–1.5mm
in thickness
Cover: 25mm (W) × 60mm (L) or less, thick-
ness of 0.18mm or less
Total thickness (after slide and cover are
sealed with cement):
2mm or less (includes increase in thick-
ness due to cement)
Do not use opaque samples.
Before Scanning
1. Hold the FH-G1 with the Nikon logo facing
up. Gently slip the prepared slide, with the
cover uppermost, into the slide insertion slot
(Figure B).
2. Adjust the position of the prepared slide
until the portion you wish to scan is within
the scan area (
5
in Figure A, indicated by the
scanning guide marks
4
).
Caution
After positioning the prepared slide, handle
the holder carefully to prevent the slide from
slipping. Although the slide is held in place
by the retaining clips inside the slot, jolting
the holder could cause the slide to slip out of
position.
3. Insert the MA-20 or MA-20 (S) S
LIDE MOUNT
ADAPTER into the scanner and check that
both the scanner and computer are on.
4. Holding the FH-G1 with the Nikon logo
facing up, slide the holder into the MA-20 or
MA-20 (S) film slot in the direction indicated
by the arrow (Figure C).
Caution
Do not turn the scanner on with the holder
inserted in the film slot. Turning the scanner
on with media inserted in the adapter will
cause an error during initialization.
Making a Scan
Once inserted in the scanner’s film slot, slides
glass (for microscope) can be scanned in the
same way you would scan a mounted film
positive. Consult the Nikon Scan software
manual for details.
Caution
Scans performed using the Digital ICE cubed
(Digital ICE
3
) function may exhibit noise. Do
not use Digital ICE cubed (Digital ICE
3
)
when scanning slides glass (for microscope).
The lower edge of the aperture may be
visible in the scan. Because the edge of the
aperture is positioned lower than the pre-
pared slide, it may be out of focus, but this
does not indicate a malfunction.
After Scanning
After completing the scan, reverse the order of
the steps listed above under “Before Scanning”
to remove the holder from the scanner and take
the prepared slide from the holder.
Caution
You may find that the slide sticks slightly
when removed from the holder. This is
normal, and the slide can be removed from
the holder by slipping it out slowly.
Maintenance
Should the product become dirty, clean it gently
with a soft cloth.
Do not expose the product to volatile chemi-
cals such as alcohol, thinner, or benzine, as this
could cause discoloration or other deteriora-
tion.
Specifications
Effective scan area:
22.9 mm × 35 mm (area indicated by
scanning guide marks)
External dimensions:
53 mm (W) × 106 mm (L) × 7 mm
(H)
Weight: approximately: 20g
1
2
3
4
5
6
7
Parts List
1
Slot cover (with four retaining clips
inside)
2
Slide insertion slot
3
Aperture
4
Scanning guide marks
5
Scan area
6
Holder entrance slide insertion
guides (slides too thick to pass
through guides can not be used)
7
Direction of insertion (arrow)
Adapter
1
Einschubabdeckung (mit vier
inneren Montageclips)
2
Präparataufnahme
3
Scanmaske mit Öffnung
4
Führungsmarkierungen
5
Scanbereich
6
Präparatführung (Präparate, die
nicht durch die Führung passen,
können mit dem Adapter nicht
verwendet werden)
7
Einschubrichtung (Pfeil)
Onderdelenlijst
1
Afdekking (met 4 bevestigingsclips
aan de binnenkant)
2
Inbrenggleuf objectglas
3
Scandiafragma
4
Scangeleidemarkeringen
5
Scangebied
6
Inbrenggeleiders objectglas
(objectglazen die te dik zijn voor de
geleiders passen niet in de houder)
7
Inbrengrichting (pijl)
Lista delle parti
1
Copertura della scanalatura (con
quattro clip di ritegno all'interno)
2
Apertura d’inserimento del vetrino
3
Apertura di scansione
4
Tacche guida per la scansione
5
Area di scansione
6
Guide d’inserimento del vetrino all’
entrata del porta vetrini (non si
possono usare vetrini troppo spessi
che non passano attraverso queste
guide)
7
Direzione d’inserimento (freccia)
Liste des pièces
1
Protecteur du récepteur porte-lame
2
Fente d’insertion des diapositives
3
Ouverture de numérisation
4
Repères de numérisation
5
Zone de numérisation
6
Repères d’insertion de la diapositive
dans le porte-objet (les diapositives
trop épaisses pour passer entre ces
repères ne peuvent pas être utilisées)
7
Direction d’insertion (flèche)
Listado de Partes
1
Tapa de la abertura (con cuatro
clips de sujeción interior)
2
Ranura de introducción de
diapositivas
3
Abertura del escáner
4
Marcas de señalización para escanear
5
Area para escanear
6
Señaladores para la introducción de
diapositivas que se encuentra en la
entrada del portaobjetos (no se
pueden utilizar diapositivas que sean
demasiado gruesas como para pasar
por los señaladores)
7
Dirección de la introducción (flecha)
Seitenansicht 0
1 2

Inhaltsverzeichnis

Seite 1 - English Deutsch

FH-G1ItNlEsFrDeEnJpMedical Slide Holder FH-G1 User’s ManualAdapter für medizinische Präparate FH-G1 BedienungsanleitungPorte-Objet Médical FH-G1 Ma

Seite 2 - Español ItalianoNederlands

Félicitations pour l’achat de votre porte-objetmédical FH-G1.Le FH-G1 est spécifiquement conçu pour la numé-risation d' objet médical avec les sc

Kommentare zu diesen Handbüchern

Keine Kommentare